TRAUMA RECON SYSTEM (TRS)
TROUBLESHOOTING GUIDE
Note: Drive unit consists of a handpiece, power module, and lid.
Problem Possible Causes Remedy
Drive unit does not start. No power module in handpiece. Insert charged power module.
Power module is drained. Charge power module.
Mode switch is set to LOCK. Set mode switch to DRILL/REAM,
SAW, or OSC DRILL.
Energy saving feature activated and Set mode switch to LOCK and then
drive unit automatically turned off. back to the applicable operating mode.
Power module is inoperable (service Send power module to
indicator lamp may be lit when DePuy Synthes Power Tools
information button is pressed). Service Department.
Trigger was depressed before power Wait 1-2 seconds when switching
module registered. from lock mode to working mode
before depressing the trigger.
Drive unit lacks power. Power module is drained. Charge power module.
Incorrect attachment used (e.g. drilling Change attachment.
speed vs. reaming speed attachment).
Mode switch set to incorrect mode. Set to correct mode.
Drive unit suddenly stops. Power module is drained. Charge power module.
Drive unit has overheated. Allow to cool to room temperature.
Drive unit is inoperable. Send to DePuy Synthes Power Tools
Service Department.
Drive unit continues to run after Trigger is jammed by residue. Immediately turn mode switch to LOCK.
releasing trigger. Clean and oil trigger according to care
and maintenance guidelines.
Drive unit or attachment becomes Drive unit or attachment is used Allow drive unit or attachment to cool.
excessively hot. too long. (see Duty cycles section)
Saw blade is not fi rmly tightened. Tighten saw blade.
DRIVE UNITS AND ATTACHMENTS
Set to correct mode.
Change attachment.
Drive unit is too slow.
Mode switch set to incorrect mode.
Incorrect attachment used (e.g. reaming
speed vs. drilling speed attachment).
Seitenansicht 0
1 2 3 4 5 6 7

Inhaltsverzeichnis

Seite 1 - TROUBLESHOOTING GUIDE

TRAUMA RECON SYSTEM (TRS)TROUBLESHOOTING GUIDE Note: Drive unit consists of a handpiece, power module, and lid. Problem Possible Causes Remedy Drive

Seite 2 - DRIVE UNITS AND ATTACHMENTS

Troubleshooting Guide Problem Possible Causes Remedy Drive unit saws too fast or aggressively. Mode switch set to incorrect mode. Set to SAW mode.

Seite 3 - ATTACHMENTS AND INSTRUMENTS

Problem Possible Causes Remedy Diffi culty inserting instruments into Attachment or instrument excessively Replace attachment or instrument. att

Seite 4 - POWER MODULE

Troubleshooting Guide Problem Possible Causes Remedy Power module cannot be inserted Power module inserted in wrong Turn power module to align

Seite 5 - UNIVERSAL BATTERY CHARGER II

Troubleshooting Guide Problem Possible Causes Remedy ON/OFF display does not illuminate. Charger is switched off. Turn charger power switch on. P

Seite 6

Problem Possible Causes Remedy Power module/battery is Power module/battery is being inserted Insert power module/battery into impossible to ins

Seite 7

Synthes (Canada) Ltd. 2566 Meadowpine Boulevard Mississauga, Ontario L5N 6P9 Telephone: (905) 567-0440 To order: (800) 668-1119 Fax: (905) 567-3185Sy

Kommentare zu diesen Handbüchern

Keine Kommentare